The PP-80 is a passive, low-impedance attenuation probe that is designed and calibrated for use with instruments having an input impedance of 1M ohm paralleled by 20pF; however, it may be recalibrated for use with instruments having an input capacitance of 15 to 40pF. The probe incorporates a two position slide switch in the head which selects attenuation of x1 or x10.
Features:
- BNC to IC Test Hook, with selectable X1 (Bandwidth DC to 6MHz), X10 (Bandwidth DC to 60MHz) switch
- Includes probe, spring hook, ground lead, IC test tip, adjusting tool, adjustment tool, and channel identifier clips
- The probe incorporates a two position slide switch in the head which selects attenuation of x1 or x10
- Designed and calibrated for use with instruments having an input impedance of 1M ohm paralleled by 20pF; however, it may be recalibrated for use with instruments having an input capacitance of 15 to 40pF
- Meets EN61010-031 CAT II safety requirements
Specifications:
- Input Capacity:
- X1: 85pF-115pF
- X10: 18.5pF-22.5pF
- Oscilloscope Input Capacity: 15pF-40pF
- Input Resistance:
- X1: 1MΩ ±2%
- X10: 10MΩ ±2%
- Bandwidth:
- X1: DC 6MHz
- X10: DC 60MHz
- Max Input Voltage:
- X1: <200V DC +Peak AC
- X10: <600V DC +Peak AC
Includes:
- Probe
- Spring hook
- Ground lead
- IC test tip
- Adjusting tool
- Channel identifier clips